Biovantion Inc.

Concentrate on IVD reagents

Manufacturer from China
Verified Supplier
4 Years
Home / Products / Elisa Test Kit /

EBV-VCA IgA Ab Test ELISA Kit Enzyme Immunoassay Test Plasma Specimen

Contact Now
Biovantion Inc.
Visit Website
City:beijing
Province/State:beijing
Country/Region:china
Contact Person:MrSteven
Contact Now

EBV-VCA IgA Ab Test ELISA Kit Enzyme Immunoassay Test Plasma Specimen

Ask Latest Price
Video Channel
Brand Name :BIOVANTION
Model Number :BE501A
Certification :ISO13485
Place of Origin :CHINA
MOQ :1 Box
Payment Terms :L/C, Western Union, T/T
Supply Ability :50000 Box/Month
Delivery Time :5-8 working days
Price :Negotiable
Packaging Details :Carton
Speciman :Serum or Plasma
Read Time :60 Mimutes
Storage :2-8℃
EXP :18 Months
Size :96 Test/Kit
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

EBV-VCA IgA Ab ELISA Kit Catalog No.: BE501A
Enzyme immunoassay for the detection of IgA antibody to the viral capsid antigen (VCA) of Epstein-Barr Virus (EBV) .


1. PRINCIPLE OF THE ASSAY
The EBV-VCA IgA Ab ELISA Kit utilizes a detection system where microplate wells are coated with recombinant antigen corresponding to a highly antigenic segment EBV-VCA. Serum or plasma specimens, controls are added to the wells. During incubation, IgA antibodies specific for EBV-VCA present in the specimen will bind to the recombinant antigen fixed onto the microplate wells. The wells are washed to remove unbound materials, and anti-human IgA peroxidase conjugate will bind to the antigen-antibody complex and excess unbound enzyme conjugates are again removed by washing. The enzyme substrate, tetramethylbenzidine (TMB), is added upon incubation the substrate will be hydrolyzed by the bound enzyme and a blue or blue-green colour develops in wells containing EBV-VCA IgA antibodies. The enzyme reaction is stopped by the addition of sulphuric acid. The intensity of colour developed is read spectrophotometrically at 450nm (450 nm/630 nm) and is proportional to the amount of antibodies present in the specimen.
REAGENTS

2. Materials provided with the kits:

ItemDescription96T480T
1Microtiter Well15
2Negative Control1ml5ml
3Positive Control1ml5ml
4Sample Dilution12ml60ml
5Enzyme Conjugate12ml60ml
6Wash Buffer Concentrate (20x)30ml150ml
7Substrate Solution A6ml30ml
8Substrate Solution B6ml30ml
9Stop Solution6ml30ml
10Plastic Bag15
11Seal Paper315
12Manual1 copy5 copy


3. ASSAY PROCEDURE

  1. Bring ELISA Kit (all reagents), and Specimens to room temperature before use (approximately 30 minutes).
  2. Dilute concentrated wash buffer 1:19 with ddH2O. For example, 5 ml of wash buffer concentrate should be diluted to a total volume of 100 mL with deionized or distilled water.
  3. For each test, set one blank well as background control, two positive and three negative controls. Dispense 100ml positive control and negative control duplicate into individual wells. Neither samples nor HRP-Conjugate should be added into the Blank well.
  4. Dispense 100ml of Sample Dilution into individual test wells except the controls.
  5. Add 10ml of each test sample into the test wells; vortex to mix.
  6. Incubate for 30 minutes at 37°C
  7. Wash each well 5 times by filling each well with diluted wash buffer, then inverting the plate vigorously to get all water out and blocking the rim of wells on absorbent paper for a few seconds.
  8. Add 100ml of Enzyme Conjugate to each well. Mix it gently by swirling the microtiter plate on flat bench for 1 minutes. Do not add Enzyme Conjugate to the blank well.
  9. Incubate for 20 minutes at 37°C
  10. Wash the plate 5 times as step 7.
  11. Add one drop (50ml) of Substrate Solution A (HRP-substrate) to each well, then add one drop (50ml) of Substrate Solution B (TMB) to each well. Mix gently and incubate at 37°C for 30 minutes. .
  12. Add one drop (50ml) of Stop Solution to each well to stop the color reaction. Read the OD value at 450 nm/630 nm with dual filter plate reader. It is option to read the OD value at 450 nm with single filter plate reader. (using the OD value of the blank well to correct all the OD reading from all wells)

EBV-VCA IgA Ab Test ELISA Kit Enzyme Immunoassay Test Plasma SpecimenEBV-VCA IgA Ab Test ELISA Kit Enzyme Immunoassay Test Plasma Specimen